Honeywell Midas-M Silane (SiH4) Sensor, 0-20 ppm, 2 years
The Midas-M Sensor Cartridge enables the detection of a wide range of asphyxiant, toxic, flammable, pyrophoric, corrosive, and oxidizer gases. The plug & play Sensor Cartridge is located inside the detector’s chassis.
Gas Measured | Silane (SiH4) | |
---|---|---|
Cartridge Part Number | MMS-X2 | |
Sensor Technology | 3 electrode electrochemical cell | |
Measuring Range | SiH4 | 0 ppm to 20 ppm |
Default Alarm 1 | 2.5 ppm (rising) | |
Default Alarm 2 | 5 ppm (rising) | |
Accuracy | <±5% of measured value | |
Exposure to SiH4 10 ppm for 5 minutes | ||
Response Time (t62.5) | Typical 5 seconds | |
Sensor Cartridge Life Expectancy | 24 months under typical application conditions | |
Operating Temperature | 0ºC to 40ºC (32ºF to 104ºF) | |
Effect of Temperature | Zero | <±0.04 ppm/ºC |
Sensitivity | <±0.8% of measured value/ºC | |
Operating Humidity (continuous) | 10% RH to 90% RH | |
Effect of Humidity | Zero | <±0.005 ppm/% RH |
Sensitivity | <±2% of measured value/% RH | |
Operating Pressure | 90 kPa to 110 kPa | |
Effect of Position | No effect in typical application | |
Long Term Drift | Zero | <±2% ppm/year |
Sensitivity | <10% of measured value/year | |
Calibration Gas | Silane (2.5 ppm to 7.5 ppm, default 10 ppm) | |
Challenge Gas (Bump Test) | Silane (10 ppm) | |
Warm Up Time | < 20 minutes | |
Storage Temperature | 5°C to 25°C (41°F to 77°F) | |
Other Detectable Gases | The following additional gases can be detected with this sensor cartridge. Sensor performance and characteristics will be representative of the data as tabulated above. Consult the Technical Manual to set up the Midas®-M transmitter with the designated identification code for each of the following gas types: | |
Detectable Gas | Chemical Formula | Measuring Range |
Disilane | Si2H6 | 0 ppm to 20 ppm |
Each Midas-M sensor is potentially cross sensitive to other gases and this may cause a gas reading when exposed to other gases than those originally designated. The table below presents typical readings that will be observed when a new sensor cartridge is exposed to the cross sensitive gas (or a mixture of gases containing the cross sensitive species).
Gas / Vapor | Chemical Formula | Concentration applied (ppm) | Reading (ppm SiH4) |
---|---|---|---|
Ammonia | NH3 | 100 | 0 |
Arsine | AsH3 | 1 | 0.54 |
Carbon Monoxide | CO | 2000 | 0.56 |
Chlorine | Cl2 | 5 | -0.86 |
Diborane | B2H6 | 1 | 0.75 |
Ethanol | C2H5OH | 500 | 0 |
Hydrogen | H2 | 5000 | 0.45 |
Hydrogen Chloride | HCI | 12.5 | 2.1 |
Hydrogen Fluoride | HF | 10 | 0 |
Hydrogen Sulfide | H2S | 10 | 2.7 |
Iso Propanol | C3H7OH | 2000 | 0 |
Nitrogen Dioxide | NO2 | 6 | -0.89 |
Phosphine | PH3 | 1 | 1.13 |
Sulfur Dioxide | SO2 | 50 | 3.5 |