Honeywell MIDAS Sensor Cartridge for Chlorine (Cl2) 0.18-2 ppm, Fluorine (F2) 0.18-2 ppm

$400.00

Description

Honeywell MIDAS Sensor Cartridge for Chlorine (Cl2) 0.18-2 ppm, Fluorine (F2) 0.18-2 ppm

Several Honeywell MIDAS® cartridges can be configured to read a number of different gases based on a common chemical family. Actual details for the sensor detection range are included in the sensor cartridge specification sheet and the MIDAS® technical manual. MIDAS® sensor cartridges can be ordered as part of a Kit with the universal MIDAS® transmitter, or as a replacement part. The detectable range is listed. With the exception of O2 (where ambient concentration is typically 20.9%), the sensors will display zero when no gas is present. All Cartridges with the format MIDAS-E-XXX are supplied with a 2 year (24 month) warranty.
Gas MeasuredChlorine (CI2)
Cartridge Part NumberMIDAS-E-HAL2 year extended warranty
Sensor Technology3 electrode electrochemical cell
Measuring Range (ppm)Cl20 – 2 ppm
Minimum Alarm 1 Set Point0.25 ppm
Repeatability< ± 5% of measured value
Linearity< ± 3% of measured value
Response Time t62.5≤ 5 seconds
Sensor Cartridge Life Expectancy≥ 24 months under typical application conditions
Operating Temperature0°C to +40°C (32°F to 104°F)
Effect of TemperatureNo effect (0°C to 20°C)
Zero< ± 0.005 ppm / °C (20°C to 40°C)
Sensitivity< ± 1.8% of measured value / °C
Operating Humidity (continuous)20 – 75% RH
Effect of HumidityZero< 0.003 ppm / % rH
Sensitivity< ± 0.3% of measured value / % rH
Operating Pressure70 – 110kPa
Effect of PositionNo effect in typical application
Long Term DriftZero< ± 2 ppm / year
Sensitivity< 10% of measured value / year
Calibration GasChlorine (Cl2)
Challenge Gas (Bump Test)Chlorine (Cl2)
Warm Up Time< 10 minutes
Storage Temperature+5°C to +25°C (+41°F to +77°F)
Other Detectable GasesThe following additional gases can be detected with this sensor cartridge. Sensor performance and characteristics will be representative of the data as tabulated above. Consult the Technical Manual to set up the Midas transmitter with the designated identification code for each of the following gas types.
Detectable GasChemical FormulaMeasuring Range
FluorineF20-4 ppm
Cross Sensitivities

Each Midas sensor is potentially cross sensitive to other gases and this may cause a gas reading when exposed to other gases than those originally designated. The table below presents typical readings that will be observed when a new sensor cartridge is exposed to the cross sensitive gas (or a mixture of gases containing the cross sensitive species).
Gas / VaporChemical FormulaConcentration Applied (ppm)Reading (ppm CI2)
AmmoniaNH350-1.9
Carbon MonoxideCO200000
Hydrogen ChlorideHCl91.25
Hydrogen SulfideH2S 25-16.3
Nitrogen DioxideNO2501.25 (transient)
Sulfur DioxideSO2509.1